Liquiline Online COD Analysis System for Wastewater and Process Applications
- Small Reagent Consumption
- Separation of Toxic Waste Water and Cleaning Waste Water
- Plunger Pump Quantitative
- Double Optical Path Detection
- Lack of Liquid Detection and Alarm
- Automatic Compensation of Turbidity
- Automatic Range Switching
- Key Parameter Upload
- Friendly Human-Computer Interaction
- Support IPv4 Network
-
Specification
| Measuring range | (0~100 / 200 / 500 / 1000 / 2000) mg/L |
| Error | 20% Range:±8%; 50% Range:±5%; 80% Range:±3% |
| Repeatability | ≤3% |
| 24h low concentration drift | ±5 mg/L |
| 24h high concentration drift | ≤3% |
| Minimum Maintenance Period | ≥168 h/time |
| Digital communication | RS232、RS485、RJ45 |
| Analog communication | (4~20)mA input; (4~20)mA output |
| Power requirements | (220±22)VAC;(50±1)Hz |
| Ambient temperature | (5~40)℃ |
| Size | 300mm×420mm×240mm (W×H×D) |
| Weight | <15 Kg |
-
Introduction
COD Analyzer Liquiline System offers unparalleled online measurement of chemical oxygen demand, supporting precise monitoring of wastewater cleaning capacity and various chemical reactions in industries, laboratories, urban and rural water treatment facilities, and sewage treatment systems. Automatic calibration and cleaning ability save your budget on a large scale, along with the advanced documentation processing via remote control.
-
Features
- Low reagent consumption: Reagent consumption for a single measurement is less than 3mL.
- Separation of toxic wastewater and non-toxic water: No interference with the measurement
- Double optical path detection: Real-time calculation of absorbance, eliminating the interference of the external environment on the optical system
- Insufficient reagent alarm: Automatically alerts to an insufficient amount of reagents and samples to avoid measurement failure
- Automatic compensation of turbidity: Automatic compensation for the sample's turbidity to eliminate the measurement failure
- Automatic range switching: Automatically switches to high-range through dilution when the detection result is over the current range
- Key parameter upload: Key parameters are being uploaded by default based on the protocol requirements
- Friendly man-machine interaction: Linux system making the operation easy and user-friendly
- Expanded storage: Supporting USB and TF cards for program update, data export, and data storage expansion