DPC100X Online Particle Cleanliness Analyzer Particle Counter
6-channel (0.3μm, 0.5μm, 1.0μm, 2.5μm, 5.0μm, 10μm) simultaneous particle number output
Industrial grade Laser diode, more stable
Big flow rate for industrial applications with a pump
Real-time flow monitoring and constant flow sampling control
Wider working temperature range
Intelligent particle recognition algorithm, meet the accurate measurement under different atmospheric
The OPC-6303DS online particle counter adopts the principle of optical scattering, which can accurately detect and calculate the number of suspended particles of different particle sizes in the air per unit volume. It can output the particle count of 6 channels of 0.3μm, 0.5μm, 1.0μm, 2.5μm, 5.0μm and 10μm at the same time (the default unit is pcs/m³, can be switched to pcs/L; pcs/ft3)
Applications
Chip manufacturing
Power battery manufacturing
Food Hygiene Industry
Precision machining
Food Hygiene Industry
Working Principle
When sampled particles pass through light beam (laser), there will be light scattering phenomenon. Scattered light will be converted into electrical signal (pulse) via photoelectric transformer. The bigger particles will obtain stronger pulse signal (peak value). Through peak value and pulse value quantity concentration of particles in each size can be calculate. Thus, real-time measurement data is obtained through measuring quantity and strength of scattered light.
Specification
Principle | Light Scattering |
Number of channels | >0.3μm, >0.5μm, >1.0μm, >2.5μm, > 5.0μm, >10μm |
Counting efficiency |
50%@0.3μm 100%@≥0.5μm (TSI9306,25±2℃, 50±10%RH) |
Measurement error |
<100pcs/l: ±30pcs/L >100pcs/L: ±30% of reading Condition: 25±2°C, 50 ±10%RH. (Reference instrument TSI 9306. TSI 9306 count efficiency 50% @ 0.3 µm and 100% for particles >0.45 µm) |
Time to first reliable reading | ≤8s |
Sampling interval | 1s |
Working condition | 0℃~45℃; 0~95%RH (non-condensing) |
Storage condition | -20~60℃, 0~95%RH (non-condensing) |
Working current | ≤1A |
Communication |
RS485 (standard) RJ45 (standard) 4~20mA (standard) |
Lifetime | ≥3 years |
Sampling flow rate | 2.83L/min |
Display | 3.5-inch color screen |
Alarm light (optional) |
Alarm Mode:LED flashing light Tone: Siren sound Sound Level: 110 decibels Protection Level: IP44 Operating Mode: alarm controlled by RS485 signal |
Temperature and Humidity Accuracy |
Temperature: ±1°C Humidity: ±8 %RH |
Standard | Comply with JJF 1190-2008 |